Panasonic*Operating InstructionsInstrucciones de UsoWrist Blood Pressure MonitorMonitor de presión arteriai de muñecaModei No7№ de Modeio EW-BW10Engli
I Inserting/Replacing Batteries*Always use alkaline batteries. (2 AAA-size LR03 alkaline batteries)• When replacing batteries, always use 2 fresh alka
For Batteries• The frequency of use for batteries is approximately 300 measurements (3 times per day) with fresh Panasonic alkaline batteries (AAA-siz
I Fitting the Pressure Cuff (cont.)2. Firmly wrap and fasten using the hoop and loop fastener.• A loosely applied pressure cuff will result in high b
I Obtaining Accurate MeasurementsAlways take your blood pressure at the same time while sitting In the same position and using the same wrist.It is re
I Obtaining Accurate Measurements (cont.)Do not take measurements in the positions shown below as they wili be inaccurate.Do not bend your wrist inwar
I Measuring Your Blood PressureFiSi Dis; ifisraaamdgwHR/f1R^SiS8DIA.•m^Bi |U| !► Pul;e v_4OFF/ONS№KT1. Place the arm to which the device is attache
I Measuring Your Blood Pressure (cont.)Reading the blood pressure displayBlood pressure values for both systolic and diastolic readings are determined
Advice:In general, blood pressure measurements are lower on the second attempt because you are calmer.For people who easily get tense, measuring twice
I Movement Detection FunctionIf movement is detected while measurement is being carried out and morepressure is applied by the pressure cuff, the rnar
I Irregular Pulse FunctionIf the pulse rate varies by more than ±25 % from the average while a measurement is being taken, when measurement is complet
Panasonic Oscillometric Automatic Wrist Blood Pressure Monitor Model EW-BW10 Is a device Intended to measure systolic and diastolic blood pressure and
I Saving and Recalling DataSaving Readings• All data Is retained even when the batteries go flat or are changed.1. The M mark will flash after measure
Recalling Stored DataReadings can be recalled after being saved without having to first turn thedevice on.1 OFh 1 STA'WSYS. imH, DIA. imH:N3 вгA
I After UseStorage1. Fold the end of the pressure cuff in the direction of the arrow.2. Next, fold the pressure cuff in on itself and secure it on loo
I SpecificationsPower source DC 3 V (2 /\/WLR03 size alkaline batteries)Display Digital LCDMethod of measurementOscillometric systemMeasurement range
I Care and MaintenanceDo not attempt to disassemble, repair or modify the unit.• Doing so may cause fire or cause the unit to malfunction. It may also
I TroubleshootingDisplay Status before error Cause and solutionUOiisdisplayed.The indication disappears in the middle of the operation, (the ! 3 mark
I Troubleshooting (cont.)Symptom Possible causeBlood pressure is abnormally high or low.• The posture of the person is different each time measurement
For questions or assistance with your biood pressure monitor, cail us at 1-800-338-0552.Panasonic Corporation of North AmericaOne Panasonic Way, 1H-1S
I Table of ContentsIntroduction... 3Basics of Biood Pressure
I IntroductionThank you for purchasing the Panasonic Automatic Wrist Biood Pressure Monitor EW-BW10.Measuring your own biood pressure is an important
I Important Instructions Before UseDo not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a hea
7.8.g.this is not possible, however, try to take measurements at a specified time prior to breakfast, and before you have become active. You should re
g. Do not short-circuit batteries.h. Do not attempt to recharge the batteries inciuded with the unit.12. Piease remove the cuff if abnormai operation
Easily Check Your Blood Pressure Readings Against the JNC 7* ClassificationBlood Pressure Categories^ A0. (mmHg)OQ Q(00)£0. aooOQo1009080Stage 2 Hyper
I Diagram of the Device* Only use alkaline batteries.BodyDisplayThis product includes:• Main unit• Storage case• Batteries• Operating Instructions• Gu
Comments to this Manuals